Samson V2 (Neptune)
The new generation cooling/heating forcing system is used as laboratory equipment to validate Si components in temperature ranges of -60ºc up to +150ºc at cooling / heating envelope between 350W@-40ºc 1200W@0ºc and 1200w @ 140c. The Samson V2 system has the capability to support Rack dimensions 50 X 25 x 35 cm. Samson V2 requires standard 208-230V 50Hz or 220V 60Hz outlet only. Samson Liquid cooling requires additional chiller with flow rate range of 3.5-2.0 liter per min @15c (+/- 2c). Total cooling system efficiency is ~55%.
Features
Temperature range | -60ºc to 150ºc |
Cooling power | 350W@-40/0º, 500W@-24/0º, 1200w@/0ºc 1200w@140º |
Head size | 78 X 90 mm |
Temperature change rate | ~(hot/cold): 3ºc per second |
Condensation prevention | Air Purge & Heaters |
Umbilical hose length Remote | 2.3 meters (between control box & thermal head) |
Remote communication | I^2C / Modbus TCP/UDP Protocols |
Refrigeration | R452a – complies with CA state GWP regulation (2020) |