Phase Change Thermal Forcing System for Si Validation -60°c to 150°c
The new cooling/heating forcing system is used as laboratory equipment to validate Si components in temperature ranges of -40ºc up to +150ºc at cooling/heating envelope between 150W@-40ºc 650W@0ºc and 800w @ 100c. The Samson system can support Rack dimensions 50 X 25 x 29 cm. Samson Air cooling requires standard 208-230V 50Hz or 110-130V 60Hz outlet only. Samson Liquid Cooling requires an additional chiller with a flow rate range of 3.0-1.8 liter per minute @15c (+/- 2c). Total cooling system efficiency is ~55%.
Features of Samson
Temperature range | -60ºc to 150ºc |
Cooling power | 150W@-40/0º, 280W@-24/0º, 650w@/0ºc 800w@100º |
Head size | 78 X 86 mm |
Temperature change rate | ~(hot/cold): 3ºc per second |
Condensation prevention | Air Purge & Heaters |
Thermal safety fuse limit | 150ºc (OTCB) |
Umbilical hose length Remote | 2.3 meters (between control box & thermal head) |
Remote communication | I^2C / Modbus TCP/UDP Protocols |
Refrigeration | R452a – complies with CA state GWP regulation (2020) |